Vacuum System Pumping

Overview

The pumping system produces the high vacuum required for high precision stable isotope analysis and has been designed to provide the high pumping speeds essential for the high gas flow rates required for continuous flow applications. The vacuum system involves two steps involving both low and high vacuum.

The layout of the pumping system is shown in Figure 4-3 and Figure 4-4 and comprises of a backing pump (e.g. Rotary pump) that provides the backing for a turbo-molecular pump (nEXT240D). The hose of the backing pump connects onto the bottom of the Foreline trap. The hose of the backing pump connects onto the bottom of the Foreline trap.

Figure 4-3: Schematic diagram of the pumping system

Not using an Active Penning gauge. Inverted magnetron gauge.

Figure 4-4: 3-D view of the pumping system and analyzer housing

Labels are too small and in different style. Need to get new image without captions and add them in Capture.

Low vacuum

The rotary pump is connected to the isoprime visION Mass Spectrometer by a 2m length of flexible hose, allowing the pump to be conveniently positioned behind and below the instrument. The backing line from the rotary pump is connected to a Foreline trap filled with activated alumina.

The isolation valve allows the rotary pump to be isolated from the turbo-molecular pump, enabling the rotary pump to remain operational during the venting of the mass spectrometer. This ensures that moisture from the air is not adsorbed onto the Foreline trap, which would prolong pump-down times.

The rotary pump is fitted with an oil mist filter to prevent unwanted emission of pump oil vapour into the atmosphere.

High Vacuum

The high vacuum is provided by the turbomolecular pump. An automatic vent valve is fitted to the pump, which ensures that even under conditions of power failure, the chances of rotary pump oil entering the analyzer are minimized. A filter is fitted to the vent valve to stop unwanted particles entering the vacuum system during the venting process.

The turbo-molecular pump is cooled by fans fitted into the instrument housing. One fan is mounted directly underneath the turbo pump; this fan draws air into the case, cooling the pump. Three fans are mounted on the rear panel which extract warm air from the chassis.

All the components of the turbomolecular pump are controlled by the on-board controller which sends status and diagnostic information to the software (IonOS). The power for the pump is supplied by the main power supply unit (MPSU).

Please refer to the maintenance section and the manufacturer’s manuals for service procedures for all components of the pumping system.

Pirani Gauge

A MicroPirani Vacuum Transducer (Figure 4-5) is used to measure the pressure in the analyzer pumping backing line. This gauge is located adjacent to the turbo pump roughing port and continuously measures the vacuum whenever the electronics unit is powered.

Substitute with 3D CAD version of the gauge and its location

Related image

Figure 4-5: MKS 925 MicroPirani Vacuum Transducer

Ionization Gauge

A Cold-Cathode Vacuum Transducer (Figure 4-6) is fitted to the ion optics housing to monitor the pressure inside the analyzer. The gauge is mounted on a NW25 elbow beneath the housing at the collector end.

Substitute with 3D CAD version of the gauge and its location

Image result for MKS 971B Cold-Cathode Vacuum Transducer

Figure 4-6: MKS 971B Cold-Cathode Vacuum Transducer

This gauge is controlled and monitored from the electronics unit, which is located on the right-hand side of the cubicle.

The gauge is enabled as soon as the turbomolecular pump reaches a speed of 60%. Monitoring of the pump speed enables the system to protect the gauge against over-pressure by disabling it when the speed drops below 60%.  In normal operation any condition that results in over-pressure will also cause the ion source power supply to be switched off.

Source Inlet Valve

The Source Inlet Valve (SIV) (Figure 4-7), is fitted to allow the switching between inlets that are controlled by the main centrION unit and a GC (gas chromatography) system which has an external open split and never requires any sample dilution. This device is fitted to all systems even if the GC inlet is not part of the configuration.

The Source Inlet Valve (SIV) (Figure 4-7), is a device to enable the switching between different inlets. It is connected to the low vacuum line to allow any gases that are not being admitted to the ion source for analysis to be pumped away.

Figure 4-7: Source inlet valve (SIV) linked to the vacuum system

Improve labels of image

It is part of the vacuum system because it has a direct connection to the low vacuum line, this is provided to pump away any gases that are not being admitted to the ion source for analysis.

A shut-off valve is provided to isolate this pumping line if it is necessary to carry out maintenance on the SIV without venting the whole system.

For further details please see the maintenance section of the manual.